INSTRUCTION MANUAL
Auto-Ranging
Digital Multimeter
True RMS
Measurement
Technology
• DATA & RANGE HOLD
• LOW IMPEDANCE
• AUDIBLE CONTINUITY
• MIN / MAX / RELATIVE
• TEMPERATURE
• DIODE TEST
• CAPACITANCE &
FREQUENCY
1000V
10A
40M Ω
ESPAÑOL pg. 19
FRANÇAIS pg. 37
ENGLISH
GENERAL SPECIFICATIONS
Klein Tools MM700 is an automatically ranging true root mean square
(TRMS) digital multimeter that measures AC/DC voltage, AC/DC current,
and resistance. It can also measure temperature, capacitance, frequency,
duty-cycle, test diodes and continuity.
• Operating Altitude: 6562 ft. (2000m)
• Relative Humidity: <90% non-condensing
• Operating Temp: 14°F to 122°F (-10°C to 50°C)
• Storage Temp: -4°F to 140°F (-30°C to 60°C)
• Accuracy: Values stated at 65°F to 83°F (18°C to 28°C)
• Temp Coefficient: 0.1 x (Quoted Accuracy) per °C above
28°C or below 18°C, corrections are required when ambient
working temp is outside of Accuracy Temp range
• Dimensions: 7.02" x 3.54" x 2.10" (178.3 x 89.9 x 53.3 mm)
• Weight: 14.6 oz. (414 g)
• Calibration: Accurate for one year
• Standards: Conforms to: UL STD 61010-1, 61010-2-030,
61010-2-033.
Certified to: CSA STD C22.2 # 61010-1,
61010-2-030, 61010-2-033.
IEC EN 61010-1, 61010-2-030,
61010-2-033, 61326-1.
• Pollution degree: 2
• Accuracy: ± (% of reading + # of least significant digits)
• Drop Protection: 6.6 ft. (2m)
• Ingress Protection: IP42 (except test lead jacks, see WARNINGS)
• Safety Rating: CAT IV 600V, CAT III 1000V, Class 2,
Double insulation
CAT III: Measurement category III is applicable to test and
measuring circuits connected to the distribution part of the
building’s low-voltage MAINS installation.
CAT IV: Measurement category IV is applicable to test and
measuring circuits connected at the source of the building’s
low-voltage MAINS installation.
• Electromagnetic Environment: IEC EN 61326-1. This
equipment meets requirements for use in basic and controlled
electromagnetic environments like residential properties,
business premises, and light-industrial locations.
Specifications subject to change.
2
ELECTRICAL SPECIFICATIONS
VOLTAGE (AUTO-RANGING)
Function Range Resolution Accuracy (50–60 Hz)
4.000V 1mV ±(1.0% + 5 digits)
AC Voltage 40.00V 10mV ±(1.2% + 5 digits)
(V AC) 400.0V 100mV
±(1.5% + 5 digits)
1000V 1V
400.0mV 0.1mV ±(1.0% + 8 digits)
4.000V 1mV
DC Voltage
40.00V 10mV ±(1.0% + 3 digits)
(V DC)
400.0V 100mV
1000V 1V ±(1.2% + 3 digits)
Input Impedance: 10MΩ
Frequency Range: 50 to 400Hz
Maximum Input: 1000V AC RMS or 1000V DC
CURRENT (AUTO-RANGING)
400.0μA 0.1μA
4000μA 1μA
AC Current ±(1.5% + 3 digits)
40.00mA 10μA
(μA and mA)
400.0mA 100μA
10.00A 0.01A ±(2.0% + 5 digits)
400.0μA 0.1μA
4000μA 1μA
DC Current ±(1.0% + 3 digits)
40.00mA 10μA
(μA and mA)
400.0mA 100μA
10.00A 0.01A ±(1.5% + 3 digits)
Overload Protection: 500mA / 1000V and 10A / 1000V Fuses
Frequency Range: 50 to 400Hz
Maximum Input: µA/mA setting: 400mA AC RMS / DC
10A setting: 10A AC RMS / DC
RESISTANCE (AUTO-RANGING)
Function Resolution Accuracy
400.0Ω 0.1Ω
4.000kΩ 1Ω
±(1.5% + 5 digits)
40.00kΩ 10Ω
400.0kΩ 100Ω
4.000MΩ 1kΩ
±(2.0% + 10 digits)
40.00MΩ 10kΩ
Maximum Input: 600V DC or 600V AC RMS
3
ENGLISH
ELECTRICAL SPECIFICATIONS
CAPACITANCE (AUTO-RANGING)
Range Resolution Accuracy
40.00nF 10pF ±(5.0% + 35 digits)
400.0nF 0.1nF
4.000μF 1nF
±(3.0% + 5 digits)
40.00μF 10nF
400.0μF 100nF
4000μF 1μF ±(5.0% + 5 digits)
Maximum Input: 600V DC or 600V AC RMS
FREQUENCY (AUTO-RANGING)
9.999Hz 0.001Hz
99.99Hz 0.01Hz
999.9Hz 0.1Hz
±(1.0% + 5 digits)
9.999kHz 1Hz
99.99kHz 10Hz
500.0kHz 100Hz
Sensitivity: >8V RMS
Maximum Input: 600V DC or 600V AC RMS
DUTY CYCLE
1% to 99.9% 0.1% ±(1.2% + 2 digits)
Pulse width: 0.1 – 100ms
Frequency width: 5Hz to 10kHz
Sensitivity: >8V RMS
Maximum Input: 600V DC or 600V AC RMS
TEMPERATURE
0° to 1500°F 0.1°F / 1°F ±(2.0% + 9°F)
-18° to 816°C 0.1°C / 1°C ±(2.0% + 5°C)
4
ELECTRICAL SPECIFICATIONS
OTHER MEASUREMENT APPLICATIONS
Maximum Input: 100V RMS in Voltage setting,
600V DC or 600V AC RMS in all other settings
• Diode Test: 1.5 mA max, open circuit voltage 3.0V DC
• Continuity Check: Audible signal when resistance <50Ω,
max test current 0.35mA
• Low Impedence (Low Z): Input impedence >3kΩ
Max input 600V RMS
• Auto Power off: After ~30 minutes of inactivity
• Sampling Frequency: 3 samples per second
• Bar Graph: 41 segment, 10-15 samples per second
• Overload: "OL" indicated on display, 1000V RMS in voltage
settings, 600V RMS in all other settings
• Polarity: "-" on display indicates negative polarity
• Display: 3 ¾ digit, 4000 Count LCD, TRMS technology
WARNINGS
To ensure safe operation and service of the meter, follow these
instructions. Failure to observe these warnings can result in
severe injury or death.
• Before each use verify meter operation by measuring a known
voltage or current.
• Never use the meter on a circuit with voltages that exceed the
category based rating of this meter.
• Do not use the meter during electrical storms or in wet weather.
• Do not use the meter or test leads if they appear to be damaged.
• Use only with CAT IV rated test leads.
• Ensure meter leads are fully seated, and keep fingers away
from the metal probe contacts when making measurements.
• Do not open the meter to replace batteries while the probes
are connected.
• Use caution when working with voltages above 25V AC RMS
or 60V DC. Such voltages pose a shock hazard.
• To avoid false readings that can lead to electrical shock,
replace batteries when a low battery indicator appears.
• Do not attempt to measure resistance or continuity on a live
circuit.
• Always adhere to local and national safety codes. Use personal
protective equipment to prevent shock and arc blast injury
where hazardous live conductors are exposed.
• Meter is IP42 dust & water resistant. Following any contact
with water, thoroughly dry meter and test lead jacks prior to
subsequent use.
5
ENGLISH
SYMBOLS ON METER
AC/DC Voltage or Current Resistance (in Ohms)
Audible Continuity Diode
Capacitance Hz Frequency
% Duty-cycle Double Insulated Class II
°F/°C Temperature (Fahrenheit / Celsius) Ground
Fuse (with rating below symbol) Low Impedence
Warning or Caution
To ensure safe operation and service of this meter, follow all warnings
and instructions detailed in this manual.
Risk of Electrical Shock
Improper use of this meter can lead to risk of electrical shock. Follow
all warnings and instructions detailed in this manual.
SYMBOLS ON LCD
H Data Hold Audible Continuity
Diode Auto Ranging
AC (Alternating Current) DC (Direct Current)
Low Battery Auto Power Off
MAX Maximum Value MIN Minimum Value
°F Degrees Fahrenheit °C Degrees Celsius
M Mega (value x 106) k kilo (value x 103)
m mili (value x 10-3) μ micro (value x 10-6)
n nano (value x 10 )
-9
V Volts
A Amps F Farads
Ohms Hz Hertz (Frequency)
% Duty-Cycle REL Relative Mode
Bar Graph
6
FEATURE DETAILS
1
7 9
8 10
11
2
3 5
6 4
NOTE: There are no user-serviceable parts inside meter.
1. 4000 count LCD display 7. "RANGE" button
2. Function selector switch 8. "REL" (Relative) button
3. "10A" jack 9. "MAX/MIN" button
4. "COM" jack 10. "HOLD" button
5. "VΩ" jack 11. "SEL" (Select) button
6. mA / µA jack
7
ENGLISH
FUNCTION BUTTONS
ON/OFF
To Power ON the meter rotate the Function Selector switch 2 from
the OFF setting to any measurement setting. To Power OFF the meter
rotate the Function Selector switch 2 to the OFF setting. By default
the meter will automatically Power OFF after 30 minutes of inactivity.
Reactivate meter by pressing any button. To deactivate the automatic
Power OFF feature, power the meter ON with the SEL button 11
depressed. When automatic Power OFF is deactivated, the
symbol will not be visible in the display.
When the Function Selector switch is rotated to a measurement
setting, "LEAD" flashes on the display as a reminder to check that test
leads are inserted into the appropriate jacks.
"SEL" (SELECT) BUTTON (FOR SECONDARY FUNCTIONS)
The "SEL" button 11 activates the secondary functions for each
setting accessible by the Function Selector switch 2 . For Current,
Voltage and Low Z, it toggles between AC and DC. For the other
functions, it switches between °F and °C, between Hz and %
Duty-cycle, and between Continuity, Resistance, and Diode-test.
The default function for each application is printed on the meter in
white; the secondary functions are printed on the meter in orange.
"HOLD" (DATA HOLD / BACKLIGHT) BUTTON
Press HOLD 10 to hold the measurement on the display. Press
again to release the display to return to live measuring.
Press and hold the "HOLD" button 10 for more than one second to
turn on or off the backlight. The backlight will automatically Power…